Skip to main content Skip to footer

When Field Performance of Masonry Does Not Correlate With Lab Test Results

November 16, 2015

When Field Performance
of Masonry Does Not Correlate
With Lab Test Results
Peter R. Meijer , AIA, NCARB
Peter Meijer Architect, PC
710 N.E. 21st Avenue, Suite 200, Portland, Oregon 97232
Phone: 503-517-0283 • E-mail: peterm@pmapdx.com
S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5 M e i je r • 9 9
ABSTRACT
As a result of the hypotheses posed by past studies, a series of advanced lab tests was
conducted on brick, mortar, and previous repair materials to assist in the determination of
the failure mechanisms at a high school in Portland, Oregon. However, the results indicated
a high-quality brick, severe weathering performance characteristics, no applied coatings,
and low IRA. Whereas the brick is clearly failing on site in areas with high concentration of
water, in-field performance of the brick is in conflict with lab test results for brick performance.
This paper includes presentation of field observations, research conducted by the
author, on-site material testing, and laboratory testing, including Fourier transform infrared
spectroscopy (FTIR), freeze/thaw, and petrographic analysis.
SPEAKER
Peter R. Meijer, AIA , NCARB — Peter Meijer Architect, PC
Peter Meijer has over 30 years of professional experience with an emphasis on the
preservation and assessment of older, existing, and historic buildings. As a professional
architect with a background in scientific research, Meijer has developed his career with a
focus on the unique building sciences associated with existing and historic resources. He
has become well-regarded both regionally and nationally as an expert on the diverse issues
affecting older buildings.
1 0 0 • Me i je r S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5
S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5 M e i je r • 1 0 1
BACKGROUND1
Ulysses S. Grant High School is located
in the Grant Park neighborhood of northeast
Portland. The 10-acre campus includes
an extensive collection of educational buildings
constructed between 1923 and 1970,
including the original main building (circa
1923). Positioned within a flat, pastoral
setting of trees and parkland, the buildings
constructed in the 1920s at Grant High
School (see Figure 1) form a core group of
Classical Revival-style buildings. The buildings
exhibit a variety of character-defining
features, including a bilaterally symmetrical
U-shaped plan, a bold portico supported
by fluted Ionic columns with a broad terra
cotta frieze, ancillary entries with terra cotta
Ionic columns or pilasters and classical
entablatures, a concrete foundation wall
extending above grade, terra cotta string
course, terra cotta cornice and coping, an
interior entry with boxed beam ceilings, and
the original gymnasium with its flush Ionic
column-lined entry.
When it was completed, Grant High
School was typical of the public high schools
constructed in Portland in the pre-World War
II era. Originally designed to be added onto
over time, the main building at Grant High
School would receive a north wing in 1925
and an auditorium and south wing in 1927.
In addition to being an extensible school,
the school was also reflective of fireproof
construction through its use of a reinforced
concrete structure with brick infill.
Even before construction began, difficulties
emerged concerning the supply
of bricks for the school. The Washington
State-based Denny-Renton Clay and Coal
Company could not supply the needed
bricks for the school, and the school board
was left to contract with another company
in Montana to supply the bricks.2
OWNER CONCERN
Over the last 15 years, Portland Public
Schools (PPS) noted an accelerated degree of
masonry face spalling on the original 1923
main building—particularly
when adjacent
to concentrated water
flow areas. (See Figure
2). There appeared to
be a direct correlation
between spalling of
the existing brick and
areas of potential water
intrusion. Such areas
included brick beneath
leader boxes and downspouts,
parapet walls,
areas near or beneath
When Field Performance
of Masonry Does Not Correlate
With Lab Test Results
Figure 1 – Grant High School
main portico.
Figure 2 – Typical nearsurface
spall.
the front portico support, and near roof-to
wall-connections. Other areas of spalling
were not as obvious, including the protected
south return wall near the south entry of the
main east façade.
In addition to the original 1923 main
building, similar near-surface spalls were
occurring on the old gym (circa 1923),
north wing (circa 1925), south wing (circa
1927), and auditorium building (circa 1927).
Likewise, the masonry units of the circa-
1952 addition to the north wing, new gymnasium
(circa 1956), library wing (circa 1959),
and the science building (circa 1966) were
free of spalling (Figure 3).
In general, maintenance efforts concentrated
on the repair of parapets, masonry
beneath downspouts, wall-to-roof intersections,
the portico entry roof, and other areas of direct visual correlation
between high concentrations of water run-off and spalling.
Previous repairs replaced long sections of masonry parapet, including
attempts at replacement with exterior-grade wallboard. Other
repairs to the brick spalls included patching the masonry faces with
synthetic mortar products. When these potential water intrusion
failures were repaired, the spalling did not stop. Repointing repairs
have also failed in various areas, exposing the original lime-based
bedding mortar. In addition, several isolated units of masonry in protected
elevations away from weathering forces were also spalling. The
isolated units had no direct correlation to each other, nor were the
masonry units concentrated on specific wall elevations. Ahead of a
major capital improvement program, PPS was interested in resolving
the masonry spalling or at least understanding its cause.
MASONRY WALL
The original circa 1922 Grant High School construction drawings
indicated a masonry wall between the concrete frame consisting of
three wythes of brick laid in a common bond approximately 13 inches
in depth, a 2-in. air gap, and a 2-in. interior finish face notated as tile.
The bricks are typical unit sizes based on the era and are approximately
2¼ in. high x 8½ in. long x 311/12 in. deep with ½-in. bed joints.
The concrete wall terminated at the bottom of the concrete roof slab.
The masonry wall extended as a parapet approximately 3 ft. beyond
the roof slab as a 13-in. solid masonry wall capped by a terra cotta
coping stone.
Over the last 90 years, the original lime-based mortar has been
replaced with flush-struck cement mortar. More recent repair mortars
used for repointing are readily visible as a result of the use of different
cement mortar colors. There is visual evidence of hard mortar
repointing and subsequent damage to the brick edges during curing.
FIELD INVESTIGATION
Visual observations of all four elevations indicated equal dispersion
of similar spalling around the building, leading to a conclusion
that wind-driven rain and weather conditions were not a direct influence.
In order to investigate wall areas and abstract materials that
had not been previously damaged, a section of the east elevation
was chosen that was more protected from the environment and relatively
free of unit spalling. The wall area chosen for invasive testing
1 0 2 • Me i je r S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5
Figure 3 – Gymnasium building (c. 1925).
Figure 4 – Sample wall surface: RILEM test.
and material removal had
a clear height of approximately
140 courses of brick
unbroken by window openings
or other decorative
shelf angles. Mortar beds
were consistent in period
of repointing, degree of
weathering, and color.
The wall-let confirmed
the assembly of a multiwythe
masonry wall constructed
in a typical fully
bedded bond course with
interlocking headers and
no cavities between the
first three brick courses.
A hand-held metal detector3
picked up the location
of hook-shaped, 3/32-in.-
gauge, steel-wire masonry
ties in alternating courses,
approximately 12 inches on center ties,
found to be in good condition with no deterioration.
The absence of corrosion on the
in-place brick wire ties indicated that little
moisture was present inside the multiwythe
wall. Bricks were well-bonded to each other
on all surfaces. (See Figure 4.)
Several material samples were removed,
including nine masonry units, four inches
of face-pointing mortar, four inches of bedding
mortar, two spalled ¼-in.-thick but
intact brick faces, and 4 inches of previous
patch material. A 4-in.-diameter diamondtip
circular saw and 8-in.-long helical drill
bit were used to remove the brick. To
evaluate samples free of observable defects
against failed material, additional samples
accessible from the ground were removed
from two discrete locations adjacent to the
primary sample façade.
Given the clear visual association of failure
mechanism and high concentration of
water, we cannot overlook the freeze/thaw
dynamic. The characteristic of the masonry
spalls, such as shallow depth (¼ in.), flacking
in layers, and seasonal rate increases,
are typically associated with freeze/thaw
failure. Given Portland’s climate, however,
freeze/thaw failure of masonry is relatively
uncommon.
A PPS-commissioned study in 20114
hypothesized that the failure of the brick
was potentially due to a number of separate
or cumulative conditions, including 1)
subfluorescence expansion of salts in the
masonry, 2) freeze/thaw, and/or 3) low
quality of the original 1923 brick.
As a result of these hypotheses and field
observations, it was prudent to conduct a
series of lab tests to the brick, mortar, and
patch materials to assist in the determination
of 1) the quality of the brick, 2) the physical
composition of the brick, 3) the quantity
of naturally occurring compounds in the
masonry and mortar (particularly salts in the
masonry), and 4) the quality of the mortar.
The findings would help narrow the potential
cause of the spalling and lead to a more
focused repair and maintenance process.
MATERIAL SAMPLES
Brick units were removed for testing
from two areas of the primary east façade:
the south wall of the southern entrance,
and the north wall of the northern entrance.
Bricks were removed for absorption and
freeze/thaw testing and petrographic analysis.
Both pointing and bedding mortar samples,
as well as the previous patching material,
were removed and sent to Willamette
Geological Service for petrographic characterization
and physical composition of the
materials. Two samples of the spalled face of
the brick and one sample of patching material
were sent to Chemoptix Microanalysis to
determine if sealants were used on the brick
and, if present, to determine the sealant
chemical makeup. (See Figure 5.)
TESTING AND RESULTS
In addition to laboratory testing, field
tests for preliminary absorption rates of the
masonry can be conducted following RILEM
protocols. RILEM works specifically with
measuring properties, performance, and
durability of various building materials.
During material removal, four RILEM5
tube tests were performed on masonry units
to analyze the initial rate of absorption of
the brick. The tests found that three out of
four bricks exhibited no water absorption.
The brick tested on its spalled surface had
an absorption rate of 7/8 mL per 15 minutes.
Similar field results with minimum
to no water absorption typically indicate
the presence of a surface sealer, since all
natural clay materials absorb water under
hydrostatic pressure. The presence of a surface
coating may lead to retention of water
within the brick, and thus prevent natural
capillary flow, natural drying, and water
evaporation. (See Figure 6.)
Samples sent to the lab for coating
assessment were analyzed via episcopic light
microscopy and Fourier transform infrared
spectroscopy (FTIR) per ASTM D1245 and
ASTM E1252. The results found no hydrocarbon
or organic formulations used on the
surface of the brick, but did find silica as
surface-binding material. The presence of
silica appears to be related to the initial firing
process and normal secondary recrystallization
of the brick. The fracture planes of
the brick suggest that the silica may have
recrystallized and hardened by subsequent
atmospheric exposure.
The diagrams of the brick samples in the
lab report show agglomerates distributed
S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5 M e i je r • 1 0 3
Figure 5 – Potential masonry coating.
through brick, a course-grained volcanic rock.
Lab tests showed the patching material to be composed of quartz
and calcite over the outer surface and a quartz and calcite-free interior.
The patching material’s FTIR spectragraph has a spike at 1500
cm-1 Wave numbers, which is a contribution from calcium carbonate.
PETROGRAPHIC ANALYSIS
A petrographic analysis was performed in general conformance
with ASTM C856 and ASTM C1324 (masonry mortar), and included
petrographic analysis, chemical analyses, x-ray diffraction, and thermogravimetric
analysis. First, each sample received visual examination,
after which thin, polished, blue-dyed sections were analyzed
under a polarized light microscope for information such as materials
ratio and presence or absence of different deterioration mechanisms.
These tests were used to assess the overall quality of material and
determine the causes of its deterioration, including freeze/thaw damage,
efflorescence, excessive retempering, cracking, ettringite formation,
and alkali-silica reactivity.
Following modified ASTM standards, due to a limited number of
physical samples, 24-hour immersion and 5-hour boil absorption
tests were performed on the brick. The bricks have a very low percent
of total absorption, at 9.5% for the 5-hour boil and 7.5% for the
24-hour test. The maximum saturation coefficent is 0.79, which is
0.01 over the maximum requirements for Severe Weathering bricks
recommended for the Portland climate (ASTM C216-07a, Table 1).
The initial rate of absorption (IRA) is 5.7g/min/30in2, which equates
to a very low-suction brick or brick with low initial rates of absorption.
The IRA lab tests correlated with field RILEM tube field test
results. The freeze/thaw durability tests resulted in passing performance
and, therefore, refuted the theory that below-freezing winter
temperatures were the cause of masonry spalling.
The petrographic characterization resulted in the most unusual
findings and the most relevant results related to potential failure
mechanisms. The bricks tested showed very small rounded voids
and planer voids. Performance of brick in the field is a result of both
material properties and resistance to microclimates within the brick’s
capillary void structure. Studies have shown a connection between
small voids in the material property and susceptibility to internal
freeze/thaw. Both rounded voids and planer voids are a result of the
firing and manufacturing process and are not always associated with
spalling. With natural absorption properties,
the brick is taking in a small quantity
of water in very small pores. 24-hour
immersion results are very low (7.5%).
Publication of more in-depth studies correlates
maximum saturation values for brick
with low 24-hour immersion values. The
effect of low immersion values and small
quantities of absorbed water may increase
the susceptibility in brick with smallpore
structure to freeze/thaw failure. (See
Figure 7.)
The presence of small-pore structure
and low-immersion values, combined with
a potential cleavage plane, are likely the
reasons the Grant High School brick are
spalling. Brick with smaller pores are less
capable of absorbing the expansive forces
1 0 4 • Me i je r S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5
Figure 6 – RILEM test.
Figure 7 – Micropore structure.
of freezing water. Interlaced pores,
creating linear plains parallel with
the face of the brick, create stress
failure points. (See Figure 8.)
CONCLUSION
Field observations of masonry
failures generally correspond with
known failure mechanisms. However,
it is not unusual that further
analysis reveals that in-field performance
tests are in conflict
with more refined laboratory test
results for brick performance. In
the case of Grant High School,
masonry units were clearly failing
on-site in areas with high concentration
of water, but the lab testing
indicated a very high-quality
brick with ASTM classification ratings
for Severe Weathering performance
characteristics and low
initial rates of absorption.
Upon closer visual examination,
it was observed that individual
units were failing in isolated
protected areas of the wall surface.
Failures in such areas could not
be accounted for under direct correlation
of heavy water intrusion and typical failure
mechanisms. As a result, additional
specialized laboratory testing was needed.
Under these conditions, test methods must
include FTIR and petrographic analysis.
FTIR analysis tests for maintenance processes
and applied-coating-induced deterioration.
Petrographic analysis provides
microscopic analysis of inherent material
composition and manufacturing-induced
flaws. Both tests provide more performance
data to assess the difference between infield
and lab performance.
Since the characteristics of the brick
resulted from the firing and manufacturing
process, the brick will remain susceptible to
microclimate freeze/thaw effects. The best
corrective action is to minimize the amount
of surface water with functional rain leaders
and downspouts, proper flashing systems,
and proper mortar joints and mortar
composition. Prevention of heavy surface
runoff water (overflowing roof drains) and
internal leakage (poor flashing) will assist
in controlling and managing future damage.
Additional spalls are likely to occur in the
future due to the accumulation of expansive
forces over a long period of time. Low quantities
of water contained in water vapor—
either through atmospheric or interior environmental
conditions—do not appear to be
accumulating in sufficient quantities6 to be
a contributing factor in the brick spalls.
Liquid water trapped within the micro pore
structure and rapid temperature fluctuation
with microclimates are the major factors
in surface spalling. Since liquid water is a
contributing factor, surface coatings that
retard the migration of trapped water within
the masonry wall will not enhance the performance
of the masonry.
Replacement of the spalled bricks is recommended
over further patching. Leaving
spalled brick in place will continue to worsen
the condition over time and affect adjacent
brick. Patching material will fail in both
the short term and long term, as evidenced
by the existing patch failures.
References 1. Portland Public Schools, Historic
Building Assessment, Entrix, October
2009.
2. Oregonian, June 23, 1923.
3. Tucker Emhart Parabolt Metal Detector.
4. Visual Masonry Condition Assessment,
Morrison Hershfield, November
2011.
5. RILEM is an acronym for Reunion
Internationale des Laboratoires
d’Essals et de Recherches sur les
Materiaux et des Constructions
(International Union of Testing and
Research Laboratories for Materials
and Structures) located in Paris,
France. The RILEM test conducted
on Grant High School was performed
to determine the hydrostatic pressure
applied to the test area. This
pressure can subsequently be converted
into a velocity or wind-driven
rain speed. A cylindrical tube is
filled to the “0.0 mL” graduation,
exerting a pressure of 1139.36 Pa,
or 98.1-mph wind-driven rain. This
test is used to identify a point of
entry of moisture to measure a wall’s
resistance to wind-driven rain, the
masonry’s rate of moisture absorption,
and the application and/or
effectiveness of applied water repellents.
6. N o visible signs of corrosion on the
embedded masonry ties.
S y m p o s i u m o n B u i l d i n g E n v e l o p e T e c h n o l o g y • No v e m be r 2 0 1 5 M e i je r • 1 0 5
Figure 8 – Interconnected planar voids.